SIEMENS HEALTHCARE DIAGNOSTICS INC. ADVIA 1800; CHEMISTRY ANALYZER Back to Search Results
Model Number ADVIA 1800
Device Problem Incorrect Or Inadequate Test Results (2456)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/22/2014
Event Type  malfunction  
Event Description
Discordant chloride results were obtained on three patient samples on an advia 1800 instrument.The discordant chloride results were not reported to the physician(s).The samples were retested on an alternate instrument and those results were reported to the physician(s).There are no known reports of patient intervention or adverse health consequences due to the discordant chloride results.
 
Manufacturer Narrative
A siemens healthcare diagnostics cse (customer service engineer) was dispatched to the customer site for instrument evaluation.The cse replaced the mixor motor, cell base, seals and tubing 49 assembly.The cse ran qc successfully.The cause of the discordant chloride results is unknown.The instrument is performing within specifications.No further evaluation of the device is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVIA 1800
Type of Device
CHEMISTRY ANALYZER
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS INC.
511 benedict avenue
tarrytown NY 10591
Manufacturer (Section G)
JEOL LTD, 3-1-2 MUSASHINO AKISHIMA
registration number:3003637681
tokyo, 196- 8558
JA   196-8558
Manufacturer Contact
john nelson
511 benedict avenue
tarrytown, NY 10591
9145242530
MDR Report Key3630125
MDR Text Key16007948
Report Number2432235-2014-00177
Device Sequence Number1
Product Code CGZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K990346
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/22/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberADVIA 1800
Device Catalogue Number073-A021-01
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/22/2014
Initial Date FDA Received02/17/2014
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
-
-