A steris service technician arrived at the facility, evaluated the ot1200 table and found the rosette teeth which lock the leg holder into position would not unlock.The technician replaced the rosette locks and pins, tested the ot1200 table and confirmed the unit was operating to specification.The unit was installed in july of 2013 and is currently under steris warranty.The rosette locking mechanisms will be sent back to steris (b)(4) for further evaluation.Steris has confirmed this is an isolated event, however, will continue to monitor any reports related to the issue described.
|