• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER MEDICAL-KALAMAZOO STAIR PRO - MODEL 6252; STRETCHER, HAND-CARRIED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER MEDICAL-KALAMAZOO STAIR PRO - MODEL 6252; STRETCHER, HAND-CARRIED Back to Search Results
Catalog Number 6252000000
Device Problems Bent (1059); Device Damaged Prior to Use (2284)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/28/2014
Event Type  malfunction  
Event Description
It was reported by service report that the bottom of the stair chair frame was bent to the extent that it could not be used.No patient was affected and no adverse consequence or clinically relevant delay in treatment was reported.
 
Manufacturer Narrative
Upon completion of the investigation it was found that the optional fold down foot section weldment and support assembly was bent and inoperable.If the optional fold down foot section weldment is inoperable the standard foot rest support would still be available for patient foot support.It was also found that the front patient left caster was bent, however was still able to roll and swivel during patient transport.
 
Event Description
It was reported by service report that the bottom of the stair chair frame was bent to the extent that it could not be used.No patient was affected and no adverse consequence or clinically relevant delay in treatment was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STAIR PRO - MODEL 6252
Type of Device
STRETCHER, HAND-CARRIED
Manufacturer (Section D)
STRYKER MEDICAL-KALAMAZOO
3800 east centre avenue
portage MI 49002
Manufacturer (Section G)
STRYKER MEDICAL-KALAMAZOO
3800 east centre avenue
portage MI 49002
Manufacturer Contact
anthony noora
3800 east centre avenue
portage, MI 49002
2693292100
MDR Report Key3634981
MDR Text Key4262416
Report Number0001831750-2014-01471
Device Sequence Number1
Product Code FPP
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/28/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number6252000000
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/28/2014
Initial Date FDA Received02/19/2014
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received03/27/2014
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-