MEDTRONIC HEART VALVES ACCUTRAK DELIVERY CATHETER SYSTEM; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED
|
Back to Search Results |
|
Model Number DCS-C4-18FR |
Device Problems
Positioning Failure (1158); Device Operates Differently Than Expected (2913)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 01/29/2014 |
Event Type
malfunction
|
Event Description
|
Medtronic received information that during the implant of this transcatheter bioprosthetic valve; after rotating the micro knob several times during deployment of the valve, the capsule would not move to deploy the valve.The entire system including the dcs and valve was removed from the patient.The valve was loaded on a new catheter.Again, the capsule would not move to deploy the valve after several turns of the micro knob.The dcs and the valve were removed from the patient.A new catheter was used along with a new valve.During deployment and the micro knob still had difficulty moving the capsule for deploying the valve.The implanter used the macro knob to deploy the valve, stopping at the 2/3 mark, tension was applied to the system and the valve was fully deployed; however it dove 8 mm and was snared into an ideal position of 4mm.The patient remained stable and was reported to be doing well.
|
|
Manufacturer Narrative
|
Product analysis: the product has been returned to medtronic's quality laboratory and continues to undergo analysis.Conclusion: at the conclusion of the analysis, a follow up report will be submitted.(b)(4).
|
|
Manufacturer Narrative
|
The lot history record was reviewed and showed that this product met all manufacturing specifications for product released for distribution.No issues were identified that would have impacted this event.Upon receipt at medtronic¿s quality laboratory, the handle was intact.The micro knob (thumb wheel) retracted and advanced the capsule.The macro (cursor) moved to fully advanced and retracted positions and locked in place when released.The distal tip of the capsule seated flush against the plunger tip when fully advanced.A kink was observed on the proximal end of the capsule.Tiny abrasions and stress marks were observed along the distal tip of the capsule.Damage noted on the drive screw was consistent with damage caused during clicking of the unit.Simulation testing was completed with a 31mm valve, with no clicking noted during valve loading.During deployment, after 20 millimeters of cursor travel the unit clicked.At that point, upward pressure was applied to the cursor, resulting in complete deployment.The clicking complaint was confirmed.However, it was not reported that the user applied the technique of lifting the cursor as prescribed in the instructions for use (ifu).A conclusive cause could not be determined, as the difficulty reported by the user may be related to not following the ifu mitigation.
|
|
Search Alerts/Recalls
|
|
|