Additional narrative: according to the additional evaluation, the nail was received bent about 70mm above the tip.The relevant dimensions of the received titanium elastic nails were checked and found to be in compliance with the technical drawings and specifications.The examination of the raw-material testing certificate and the manufacturing papers showed no deviations regarding material analysis, strength and structural stability.The values were in compliance with the international standard iso.Based on the complaint description we assume that excessive strain resulting of the patients fall has led to re-fracture and the deformation of the nails.No product fault could be detected.Date reported in error.The implant was done in october 2013, but the exact date is not available.
|