DEPUY SYNTHES POWER TOOLS STANDARD CRANIOTOME, BLACK MAX; BUR, SURGICAL, GENERAL & PLASTIC SURGERY - CRANIOTOME Back to Search Results
Catalog Number B-GREEN
Device Problems Break (1069); Overheating of Device (1437)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/09/2014
Event Type  malfunction  
Event Description
It was reported that during engineering evaluation, it was discovered that the craniotome device had "a damaged neuro tip and was overheating".This event was not related to surgery.There were no injuries or medical intervention or prolonged hospitalization associated with this event.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Manufacturer Narrative
The actual sample was returned for evaluation.Reliability engineering evaluated the device and discovered the device had "a damaged neuro tip and was overheating".The assignable root cause was determined to be due to normal wear over time.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STANDARD CRANIOTOME, BLACK MAX
Type of Device
BUR, SURGICAL, GENERAL & PLASTIC SURGERY - CRANIOTOME
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
rodney crain
4500 riverside drive
palm beach gardens, FL 33410
5616271080
MDR Report Key3647900
MDR Text Key4200709
Report Number1045834-2014-11516
Device Sequence Number1
Product Code GFF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK974025
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Service and Testing Personnel
Type of Report Initial
Report Date 01/09/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberB-GREEN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/18/2013
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/09/2014
Initial Date FDA Received02/26/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/17/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-