Result: the aspiration tubing on/off switch is cracked underneath the toggle switch.The 3-way stop cock (non-penumbra device) is cracked where the luer fitting connects to the stop cock.Conclusion: the device was returned for analysis and has been evaluated.The complaint indicates that there was a leak from the on/off switch that connected to the aspiration tubing.Evaluation of the returned product confirmed a crack in the aspiration tubing.In addition, aspiration from the tubing to the canister was confirmed however, the leaks in both the aspiration tubing and crack in the non-penumbra 3-way stop cock prevented efficient aspiration into the system.The crack in the aspiration tubing is of unknown origin.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This mdr is associated with mdr 3005168196-2014-00181.
|