Catalog Number 4701000000 |
Device Problems
Break (1069); Device Slipped (1584)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 02/11/2014 |
Event Type
malfunction
|
Event Description
|
It was reported via repair work order that the footsection wont latch due to damaged foot section weldment.No patient was affected and no adverse consequence or clinically relevant delay in treatment was reported.
|
|
Manufacturer Narrative
|
Customer will order parts and install themselves.
|
|
Event Description
|
It was reported via repair work order that the footsection wont latch due to square tubes on the side of the foot section weldment being bent.No patient was affected and no adverse consequence or clinically relevant delay in treatment was reported.
|
|
Manufacturer Narrative
|
Supplemental submitted as the investigation concluded that the square tubes on the side of the foot section weldment being bent.The issue was resolved for the customer by ordering a new foot section, installing it, and verifying the bed performed to specification.
|
|