STRYKER MEDICAL-KALAMAZOO LD304 MATERNITY BED; TABLE, OBSTETRICAL, AC-POWERED (AND ACCESSORIES) Back to Search Results
Catalog Number 4701000000
Device Problems Break (1069); Device Slipped (1584)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/11/2014
Event Type  malfunction  
Event Description
It was reported via repair work order that the footsection wont latch due to damaged foot section weldment.No patient was affected and no adverse consequence or clinically relevant delay in treatment was reported.
 
Manufacturer Narrative
Customer will order parts and install themselves.
 
Event Description
It was reported via repair work order that the footsection wont latch due to square tubes on the side of the foot section weldment being bent.No patient was affected and no adverse consequence or clinically relevant delay in treatment was reported.
 
Manufacturer Narrative
Supplemental submitted as the investigation concluded that the square tubes on the side of the foot section weldment being bent.The issue was resolved for the customer by ordering a new foot section, installing it, and verifying the bed performed to specification.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LD304 MATERNITY BED
Type of Device
TABLE, OBSTETRICAL, AC-POWERED (AND ACCESSORIES)
Manufacturer (Section D)
STRYKER MEDICAL-KALAMAZOO
3800 east centre avenue
portage MI 49002
Manufacturer (Section G)
STRYKER MEDICAL-KALAMAZOO
3800 east centre avenue
portage MI 49002
Manufacturer Contact
pravin betala
3800 east centre avenue
portage, MI 49002
2693292100
MDR Report Key3658383
MDR Text Key4353736
Report Number0001831750-2014-01989
Device Sequence Number1
Product Code HDD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K950889
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/11/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number4701000000
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/11/2014
Initial Date FDA Received03/04/2014
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received04/09/2014
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-