• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC HEART VALVES DIVISION ACCUTRAK DELIVERY CATHETER SYSTEM; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC HEART VALVES DIVISION ACCUTRAK DELIVERY CATHETER SYSTEM; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED Back to Search Results
Model Number DCS-C4-18FR
Device Problems Positioning Failure (1158); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/29/2014
Event Type  malfunction  
Event Description
Medtronic received information that during the implant of this transcatheter bioprosthetic valve, the capsule of this delivery catheter system (dcs) would not retract due to clicking of the micro knob.The macro knob was pushed up with the thumb, without success in moving the capsule.The dcs was removed from the patient and another dcs of the same lot number was used with a successful implant.There was no impact to the patient associated with this event.The dcs was discarded.
 
Manufacturer Narrative
Product analysis: no product was returned, therefore no product analysis can be performed.Conclusion: without return of the product, no definitive conclusions could be drawn regarding the clinical observation.No adverse patient effects were reported.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCUTRAK DELIVERY CATHETER SYSTEM
Type of Device
AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED
Manufacturer (Section D)
MEDTRONIC HEART VALVES DIVISION
1851 e. deere ave
santa ana CA 92705
Manufacturer (Section G)
MEDTRONIC STRUCTURAL HEART
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
mike gaffney
8200 coral sea street ne
mounds view, MN 55112
7635265629
MDR Report Key3662834
MDR Text Key4198813
Report Number2025587-2014-00097
Device Sequence Number1
Product Code NPT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P130021
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Study,Health Professional,Company Representative,company representati
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/29/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/25/2014
Device Model NumberDCS-C4-18FR
Device Catalogue NumberDCS-C4-18FR
Device Lot Number0006919910
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/29/2014
Initial Date FDA Received03/06/2014
Supplement Dates Manufacturer Received01/29/2014
Supplement Dates FDA Received09/14/2017
Was Device Evaluated by Manufacturer? No
Date Device Manufactured10/25/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age00082 YR
Patient Weight64
-
-