"the unit was received from the field for evaluation.Refer to evaluation #(b)(4): hp 1 - unit primed and functioned properly on test unit test002.Could not duplicate any malfunction.Hp 2 - customer stated check handpiece.Microtip was not returned, only packaging.Note on box stated scrub nurse discarded by mistake.Could not verify the customer's complaint of failed prime.The dhr for microtip lot 14413-05 was reviewed and showed that the units met manufacturing specifications when released and there were no issues related to the complaint.(b)(4).
|