• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH SPECIALTY TRIATHLON FULLY CAPTURED DISTAL RESECTION GUIDE PER FILE K2869; INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH SPECIALTY TRIATHLON FULLY CAPTURED DISTAL RESECTION GUIDE PER FILE K2869; INSTRUMENT Back to Search Results
Catalog Number I-K2869KF00
Device Problems Difficult to Remove (1528); Sticking (1597)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/17/2014
Event Type  malfunction  
Event Description
After placing the im distal resection assembly into the femur surgeon resident (b)(6) started to pin the block to the femur using a wire driver on stryker system 6 power.While driving the pin into the bone it became stuck in the cutting block and was not able to be removed.
 
Manufacturer Narrative
A supplemental report will be submitted upon completion of the investigation.
 
Manufacturer Narrative
An event regarding pin seizure involving a specialty triathlon guide was reported.The event was confirmed.Method & results: -device evaluation and results: a material analysis indicated the pin galled in the guide hole.No material or manufacturing defects were observed on the device features evaluated.-medical records received and evaluation: records were not provided.There is no indication the event is patient related.-device history review: the reported device was manufactured and accepted into final stock with no discrepancies.-complaint history review: there have been no other events for the reported lot id.Conclusions: the investigation concluded that the pin galled and seized in the guide hole.The reported device was not properly identified.However, a review of the nc/capa data base for triathlon guides and pin seizure indicates that a capa was opened to determine root cause and effective corrective action.The capa team decided that the best corrective action was to change the design of the 1/8" headless pin.Change notice was implemented to add flutes to the bottom half of the pin.The fluted pins will allow the pin to act as a drill bit and clear any burrs or debris located on the inside surfaces of the cutting blocks/guides.The reported pin is contained within the scope of the capa.
 
Event Description
After placing the im distal resection assembly into the femur surgeon resident pgy 5 mary ragsdale started to pin the block to the femur using a wire driver on stryker system 6 power.While driving the pin into the bone it became stuck in the cutting block and was not able to be removed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPECIALTY TRIATHLON FULLY CAPTURED DISTAL RESECTION GUIDE PER FILE K2869
Type of Device
INSTRUMENT
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
keyla navedo
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key3675624
MDR Text Key4258707
Report Number0002249697-2014-00755
Device Sequence Number1
Product Code JWH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K123486
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/18/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberI-K2869KF00
Device Lot NumberMME8EE
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/27/2014
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/23/2015
Initial Date FDA Received03/12/2014
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received02/18/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/08/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-