STERIS MEXICO, S. DE R.L. DE C.V. V-PRO 1 PLUS STERILIZER Back to Search Results
Device Problem Chemical Spillage (2894)
Patient Problem Caustic/Chemical Burns (2549)
Event Date 02/17/2014
Event Type  No Answer Provided  
Event Description
The user facility reported an employee received a chemical burn on the back of her hand while removing a package from the v-pro 1 plus sterilizer.The employee was sent to the facility's health services to have her hand examined.No procedural delays/cancellations were reported.
 
Manufacturer Narrative
A steris service technician arrived at the facility, inspected the sterilizer and found the unit operating to specification.The reported event could not be duplicated.As a preventive maintenance, the technician replaced the unit's oil and filters, completed a test cycle and returned the unit to operation.The employee subject of the reported event was not wearing the appropriate ppe as recommended in the v-pro 1 plus operator manual.The operator manual (6-11) states, "unload sterilization unit: steris recommends (in accordance with ansi/aami st58, 2005) wearing chemical -resistant gloves when using the sterilization unit." in addition, the operator manual (1-2) states, "danger - chemical injury hazard: any visible liquids in the chamber must be treated as concentrated hydrogen peroxide.Observe all hydrogen peroxide handling precautions." the steris technician confirmed that the facility's protocol is to always wear gloves while unloading the v-pro 1 plus sterilizer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V-PRO 1 PLUS STERILIZER
Type of Device
STERILIZER
Manufacturer (Section D)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, nuevo leon 6719 0
MX  67190
Manufacturer (Section G)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, nuevo leon 6719 0
MX   67190
Manufacturer Contact
kathryn cadorette
5960 heisley road
mentor, OH 44060
4403927231
MDR Report Key3677669
MDR Text Key16594460
Report Number3005899764-2014-00016
Device Sequence Number1
Product Code MLR
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/13/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/17/2014
Initial Date FDA Received03/13/2014
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-