• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTUITIVE SURGICAL,INC. THORACIC GRASPER INSTRUMENT; ENDOSCOPIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTUITIVE SURGICAL,INC. THORACIC GRASPER INSTRUMENT; ENDOSCOPIC INSTRUMENT Back to Search Results
Model Number 420343-02
Device Problem Split (2537)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/18/2014
Event Type  malfunction  
Event Description
It was reported that during the reprocessing of a da vinci thoracic grasper instrument, the shaft was observed to have a slit in it.No missing or fallen pieces were reported.No patient harm, adverse outcome or injury was reported.
 
Manufacturer Narrative
The instrument was returned and evaluated.Failure analysis investigation found that the distal end of the instrument's main tube exhibited scratch marks, with sections of the tube insulation removed.The damaged areas were located approximately.80 -.30 above the wrist.The missing piece is approximately.08 x.04 in size.No other damage was found.The damage to the instrument found during failure analysis investigation does not constitute a reportable event; however, the main tube scratches with insulation removed could likely cause or contribute to an adverse event, if the malfunction were to recur.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THORACIC GRASPER INSTRUMENT
Type of Device
ENDOSCOPIC INSTRUMENT
Manufacturer (Section D)
INTUITIVE SURGICAL,INC.
sunnyvale CA
Manufacturer (Section G)
INTUITIVE SURGICAL, INC.
950 kifer rd
sunnyvale CA 9408 6
Manufacturer Contact
tabitha reed
950 kifer rd
sunnyvale, CA 94086
MDR Report Key3677815
MDR Text Key4288770
Report Number2955842-2014-01506
Device Sequence Number1
Product Code NAY
Combination Product (y/n)N
PMA/PMN Number
K050369
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation Other
Type of Report Initial
Report Date 02/19/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number420343-02
Device Lot NumberS10131031 962
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/27/2014
Is the Reporter a Health Professional? No
Event Location Hospital
Initial Date Manufacturer Received 02/19/2014
Initial Date FDA Received03/13/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/01/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-