On (b)(6) 2014, the patient was undergoing an intervention with gore® excluder® aaa endoprostheses to treat a right distal type i endoleak.After a 12 fr gore® dryseal sheath was advanced up the right side, a contralateral leg component reportedly became stuck about halfway into the sheath and would not advance further.An attempt was made to pull the delivery catheter out of the sheath, but it was reported the endoprosthesis felt like it was coming off of the delivery catheter.The device and sheath were removed together, and a new sheath was inserted to implant another device with no further issue.It was reported the difficulty advancing was caused by inserting the contralateral leg component into too small of a sheath.The patient tolerated the procedure.Evaluation summary - the contralateral leg component was returned to gore for evaluation.Visual inspection of the returned device confirmed the undeployed device was caught in the 12 fr introducer sheath.There was blood on the device, and the deployment knob had not been pulled.The introducer sheath was cut open to remove the device for further evaluation.There was no evidence of damage to the leading portion of the device or the leading olive.The trailing olive was separated from the outer shaft and slightly deformed at the trailing end.The trailing end of the endoprosthesis was pulled away from the polyimide and damaged.The evaluation of the returned device is consistent with the reported observations.The root cause(s) for the inability to insert the device through the introducer sheath was inserting a 23 mm contralateral leg component into a 12 fr sheath.The instructions for use recommends a 14 fr or larger introducer sheath for the 23 mm contralateral leg component.
|