• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK, INC. BIWIRE NITINOL HYDROPHILIC WIRE GUIDE; OCY ENDOSCOPIC GUIDEWIRE, GASTROENTEROLOGY-UROLOGY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK, INC. BIWIRE NITINOL HYDROPHILIC WIRE GUIDE; OCY ENDOSCOPIC GUIDEWIRE, GASTROENTEROLOGY-UROLOGY Back to Search Results
Catalog Number BWS-038150
Device Problem Material Separation (1562)
Patient Problem Foreign body, removal of (2365)
Event Date 02/14/2014
Event Type  Injury  
Event Description
The coating stripped off of the wire when the wire was entering the kidney through the sheath of a needle.The coating was then in the kidney.The physician was able to remove the coating from the kidney with a grasper.However, this caused the physician to lose access.According to the initial reporter, the pt did not experience any adverse effects due to this occurrence.
 
Manufacturer Narrative
Removal of device portion.Coating coming off.Still under investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIWIRE NITINOL HYDROPHILIC WIRE GUIDE
Type of Device
OCY ENDOSCOPIC GUIDEWIRE, GASTROENTEROLOGY-UROLOGY
Manufacturer (Section D)
COOK, INC.
bloomington IN 47402
Manufacturer Contact
rita harden, dir
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key3683056
MDR Text Key4254744
Report Number1820334-2014-00086
Device Sequence Number1
Product Code OCY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K082536
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 02/14/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date06/30/2016
Device Catalogue NumberBWS-038150
Device Lot Number10288143
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/13/2014
Initial Date FDA Received03/11/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/01/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-