• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER MEDICAL-KALAMAZOO ISOFLEX AFTERMRKT MATT MODEL; MATTRESS, FLOTATION THERAPY, NON-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER MEDICAL-KALAMAZOO ISOFLEX AFTERMRKT MATT MODEL; MATTRESS, FLOTATION THERAPY, NON-POWERED Back to Search Results
Catalog Number 2800100999
Device Problems Patient-Device Incompatibility (2682); Material Integrity Problem (2978)
Patient Problem Tissue Breakdown (2681)
Event Date 02/19/2014
Event Type  Injury  
Event Description
It was reported by the user facility that the patient's skin allegedly broke down while on the mattress.
 
Event Description
It was reported by the user facility that the patient's skin allegedly broke down while on the mattress.
 
Manufacturer Narrative
Though no product was returned from the user facility in (b)(6), pictures were reviewed by a quality engineer and marketing specialist and the pictures did not appear to show a product malfunction.It was explained to the customer that the covers are loose to provide proper patient immersion and pressure redistribution as part of the therapeutic properties of the mattress.If a patient¿s skin has become marked from laying on the mattress, this may indicate improper hospital turning protocols or a pre-existing skin condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ISOFLEX AFTERMRKT MATT MODEL
Type of Device
MATTRESS, FLOTATION THERAPY, NON-POWERED
Manufacturer (Section D)
STRYKER MEDICAL-KALAMAZOO
3800 east centre avenue
portage MI 49002
Manufacturer (Section G)
STRYKER MEDICAL-KALAMAZOO
3800 east centre avenue
portage MI 49002
Manufacturer Contact
anthony noora
3800 east centre avenue
portage, MI 49002
2693292100
MDR Report Key3686173
MDR Text Key17385407
Report Number0001831750-2014-02371
Device Sequence Number1
Product Code IKY
Combination Product (y/n)N
Reporter Country CodeBR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/20/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Other
Device Catalogue Number2800100999
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/20/2014
Initial Date FDA Received03/18/2014
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received07/01/2014
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-