DEPUY SYNTHES POWER TOOLS 12.6CM TAPERED ATTACH, BM; BUR, SURGICAL, GENERAL & PLASTIC SURGERY - ATTACHMENT
|
Back to Search Results |
|
Catalog Number B-TURQ-L |
Device Problem
Overheating of Device (1437)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 02/27/2014 |
Event Type
malfunction
|
Event Description
|
It was reported that during a spine decompression/fusion surgical procedure, it was discovered that the attachment device was ¿getting very hot when on and activated¿.There were no delays to the planned surgical procedure as a spare device was available for use.There was patient involvement reported.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Manufacturer Narrative
|
(b)(6).The actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
The actual device was returned for evaluation.Reliability engineering evaluated the device.The unit was tested which found the temperature was above specification.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to normal wear of component parts.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|