• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES EZGLIDE AORTIC CANNULA; CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES EZGLIDE AORTIC CANNULA; CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS Back to Search Results
Model Number EZF21A
Device Problem Device Contaminated During Manufacture or Shipping (2969)
Patient Problem No Patient Involvement (2645)
Event Date 02/25/2014
Event Type  No Answer Provided  
Event Description
It was reported by warehouse in (b)(4) that the ez glide aortic cannula "hair contamination was found inside the package during incoming inspection." this was found during (b)(4) incoming inspection and never had contact with the patient/hospital.
 
Manufacturer Narrative
The device evaluation is currently in process into root cause.
 
Manufacturer Narrative
Evaluation: a loose black fiber was found and the analysis was consistant with rayon.This report is a confirmed manufacturing defect.Edwards has iniatied a pra and capa to investigate this event.Edwards has a technical summary to discuss the product safety risk adn analysis of ongoing investigation work.Trends will continue to be monitored through the edwards quality system.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EZGLIDE AORTIC CANNULA
Type of Device
CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
EDWARDS LIFESCIENCES
12050 lone peak parkway
draper UT 84020
Manufacturer (Section G)
EDWARDS LIFESCIENCES LLC
12050 lone peak parkway
draper UT 84020
Manufacturer Contact
walt wiegand
12050 lone peak parkway
draper, UT 84020
8015655200
MDR Report Key3691370
MDR Text Key16685852
Report Number3008500478-2014-00040
Device Sequence Number1
Product Code DWF
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K123370
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Invalid Data
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/25/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date01/01/2017
Device Model NumberEZF21A
Device Lot Number59679155
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/07/2014
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Distributor Facility Aware Date02/25/2014
Device Age1 MO
Date Report to Manufacturer02/25/2014
Initial Date Manufacturer Received 02/25/2014
Initial Date FDA Received03/20/2014
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received05/23/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/29/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-