EDWARDS LIFESCIENCES PROPLEGE; CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS Back to Search Results
Model Number PR9
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Sinus Perforation (2277)
Event Date 02/26/2014
Event Type  Injury  
Event Description
As reported during use of proplege a coronary sinus was perforated.Is it the second event in this hospital with different anesthesiologist.The coronary sinus was engaged in position 2.Insertions attempts were brief.
 
Manufacturer Narrative
The device was not retained by the hospital for evaluation.At this time, there is no allegation of product malfunction but only an adverse event associated with the use of an edwards device.The device was not returned for evaluation.Injuries to the coronary sinus is listed as a potential complication in the product ifu.The ifu further notes: warning: if resistance is met, stop and re-evaluate proplege device position.The proplege device should not be advanced if resistance is felt, as doing so would cause the proplege device to bend or buckle.Aggressive advancement in an attempt to engage the ostium may result in perforation or other injury.Warning: continuously monitor the pressure at the tip of the catheter.A pressure change that indicates the catheter has entered the right ventricle should be noted.Do not advance the catheter further if the tip is in the right ventricle as perforation or other injury can occur.Warning: guidewire should only be used with fluoroscopy to avoid coronary sinus injury.Warning: aggressive advancement of the guidewire in an attempt to engage the ostium may result in perforation or other injury.No device malfunction is indicted in the report and the events reported are included in the labeling.No actions are needed at this time.For complaints/reports of injury without a product problem, it¿s important to show that the type of event is a known potential complication or adverse event, is included in the labeling/training materials, and in the information we provide to help avoid the issue.All labeling and training appropriately address the risk.Manufacturing records could not be reviewed as a lot number is unknown.Trends will continue to be monitored through the use of edwards quality systems.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROPLEGE
Type of Device
CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
EDWARDS LIFESCIENCES
12050 lone peak parkway
draper UT 84020
Manufacturer (Section G)
EDWARDS LIFESCIENCES LLC
12050 lone peak parkway
draper UT 84020
Manufacturer Contact
walt wiegand
12050 lone peak parkway
draper, UT 84020
8015655200
MDR Report Key3696520
MDR Text Key4261813
Report Number3008500478-2014-00042
Device Sequence Number1
Product Code DWF
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K120780
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 02/26/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberPR9
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date02/26/2014
Event Location Hospital
Date Report to Manufacturer02/26/2014
Initial Date Manufacturer Received 02/26/2014
Initial Date FDA Received03/24/2014
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-