(b)(4).The analysis results found that the el5ml device was returned in good visual condition.In an attempt to replicate the reported incident, the device was tested for functionality.Upon testing, the device was cycled, fed, and formed the remaining 5 clips as intended.In order to confirm the clips were within manufacturing specifications, the clips were evaluated using a tool that is designed to determine proper formation.During analysis the jaws opened and closed as intended.In addition the device locked out as intended.No conclusion could be reached as to what may have caused the reported incident.A batch/lot record review was performed and the batch met all final acceptance criteria.
|