It was reported that before an unknown procedure, the handpiece was plugged into the generator without any instrument plugged onto it and it came up with switch error, two activations detected.This was the second time the handpiece had been plugged in and it was the second time this error came up even though no instrument was attached.Another like device was used to complete the procedure.There were no adverse consequences for the patient.
|
(b)(4) - nose cone.Should the information be provided later, a supplemental medwatch will be sent the handpiece was received with the nose cone damaged and the mount was noted to be loose.It was connected to a generator, evaluated with a test instrument and resulted in "reactivate two switch activations detected" alert screen displayed by the generator.The instrument was disassembled to perform an internal electrical test that revealed a hand activation to ground short circuit condition at the cable level, affecting the functionality of the handpiece.In addition the acoustic isolator was torn and there was moisture evidence.The handpiece has a number of seals to prevent fluids from entering the housing.¿moisture evidence¿ describes a condition where water enters the handpiece cavity during the steam sterilization process.The primary path of ingress is the distal seal, caused by a reduction of compressive force on the distal seal.A possible cause of the nose cone being cracked is the sterilization method due to the heating and cooling of the sterilization cycles is a stressor.
|