During the procedure the surgeon ordered a 3-d fluoro spin with the siemens arcadis orbic, which was performed in the operating room.The images were reviewed by the surgeon and surgical intervention was altered based on the images provided by the device.It was later discovered that the positioning imputed prior to the scan being performed was selected supine rather than prone.This resulted in laterally images being reversed when reviewed by the surgeon.Concerned that the markers for right and left are small on the images produced.Questioned if the markers for right and left on the images could be enlarged to assist with verification of positioning.Questioned if an additional safety check with mandatory response could be added to confirm proper positioning has been selected prior to being able to start the scan.
|