HILL-ROM, INC. SYNERGY AIR ELITE BLOWER; BED, FLOTATION THERAPY, POWERED Back to Search Results
Model Number 6881CAP
Device Problems Inflation Problem (1310); Improper or Incorrect Procedure or Method (2017)
Patient Problems Pressure Sores (2326); Injury (2348)
Event Date 02/24/2014
Event Type  Other  
Event Description
The caregiver reported the mattress is not inflating properly in the seat section and has caused three new pressure ulcers on the pt.The mattress is located in the pt's home.This report was filed in our complaint handling system as complaint # (b)(4).The hill-rom representative spoke with the caregiver to discuss the allegation of three new sores developing.The caregiver stated that the pt developed three new stage 3 sores over the weekend due to the mattress not inflating.The pt was receiving treatment for these wounds and they are healing nicely.The hill-rom representative asked the caregiver about having the mattress on top of another mattress and she could not confirm.There have been no new sores since the repair.The reported injury is serious in nature per fda definition.The cause of the malfunction is related to misuse.Development of pressure ulcers is multifactorial and cannot be only attributed to performance of the surface.Risk factors include protein-calorie malnutrition, microclimate (skin wetness caused by sweating or incontinence), diseases that reduce blood flow to the skin, such as arteriosclerosis, or diseases that reduce the sensation in the skin, such as paralysis or neuropathy.Position changes are key to pressure are key to pressure sore prevention and treatment.These changes need to be frequent, repositioning needs to avoid stress on the skin, and body positions need to minimize the risk of pressure on vulnerable areas.
 
Manufacturer Narrative
The hill-rom technician performed the investigation and per the caregiver, the mattress was not inflating properly in the seat section causing the pt multiple pressure ulcers.Pt was given wound dressings.The caregiver would not release any other information.The hill-rom technician inspected the mattress and found the account had placed this hill-rom mattress over another mattress.The synergy air elite is a mattress replacement, not an overlay and this caused three seat section bladders in the mattress to partially disconnect and the shear protector to be pulled to one side.The technician corrected the issue by reinstalling the bladders and shear protector.The technician also replaced a clogged air filter as a preventative maintenance measure.The mattress is now installed correctly on the bed and is functioning as designed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY AIR ELITE BLOWER
Type of Device
BED, FLOTATION THERAPY, POWERED
Manufacturer (Section D)
HILL-ROM, INC.
charleston SC
Manufacturer Contact
jennifer morris
1069 state route 46 east
batesville, IN 47006
8129313121
MDR Report Key3716541
MDR Text Key4336682
Report Number1045510-2014-00001
Device Sequence Number1
Product Code IOQ
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 02/24/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number6881CAP
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/24/2014
Initial Date FDA Received03/24/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/01/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age53 YR
Patient Weight76
-
-