Upon receipt at our post market quality assurance laboratory, a thorough evaluation of the device was performed.A visual inspection of the device header and case noted no anomalies.The device was then exposed to simulated heart load conditions, and the pacing and sensing functions were tested.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.However, initial testing had determined this device did not meet longevity expectations.The titanium case was opened and visual inspection of the internal components revealed no irregularities.When connected to an external power source, the device passed all tests and operated normally.No high current drain or other device malfunction was identified during analysis.Despite exhaustive testing, the cause of the premature battery depletion could not be determined.
|