• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FRESENIUS MEDICAL CARE NORTH AMERICA FRESENIUS 2008K@HOME

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FRESENIUS MEDICAL CARE NORTH AMERICA FRESENIUS 2008K@HOME Back to Search Results
Model Number 2008K@HOME
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/29/2014
Event Type  malfunction  
Event Description
A hemodialysis patient reported during treatment a blood leak occurred.Home patient reported to fresenius canada staff that he has lost a circuit of blood (approx.300ml).He reports that he wakes up in the middle of the night and finds the venous pressure is zero and machine will not run his blood.During this incident the transducer was wet.He also stated that the venous chamber level was rising throughout the night.The 2008k home machine was checked and found to be okay.Patient had no adverse effects and required no medical intervention.Sample has not been returned to the mfr.
 
Manufacturer Narrative
The plant investigation has not yet been completed.A follow up report will be filed upon completion of the investigation.This medwatch report is associated with mdr report numbers: 8030665-2014-00220, 00221, 00222, 2937457-2014-00378, 00379 and 00380.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FRESENIUS 2008K@HOME
Manufacturer (Section D)
FRESENIUS MEDICAL CARE NORTH AMERICA
2637 shadelands dr.
walnut creek CA 94598
Manufacturer (Section G)
CONCORD PLANT
4040 nelson ave
concord CA 94520
Manufacturer Contact
jennifer nabukenya
2637 shadelands dr.
walnut creek, CA 94598
8006621237
MDR Report Key3721022
MDR Text Key16801211
Report Number2937457-2014-00378
Device Sequence Number1
Product Code ONW
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Consumer,Company Representative
Reporter Occupation Patient
Type of Report Initial
Report Date 02/10/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number2008K@HOME
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/10/2014
Initial Date FDA Received03/12/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
SINGLE NEEDLE BLOOD TUBING SET
-
-