• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: IMMUCOR, INC. CAPTURE-R SELECT; SOLID PHASE SYSTEM FOR THE IMMOBILIZATION OF HUMAN ERYTHROCYTES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

IMMUCOR, INC. CAPTURE-R SELECT; SOLID PHASE SYSTEM FOR THE IMMOBILIZATION OF HUMAN ERYTHROCYTES Back to Search Results
Catalog Number 0006446
Device Problem Incorrect Or Inadequate Test Results (2456)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/11/2014
Event Type  malfunction  
Event Description
On (b)(6) 2014, a customer reported unexpected compatible crossmatch results when using capture-r select on a galileo neo instrument on (b)(6) 2014.
 
Manufacturer Narrative
Immucor technical support used remote electronic access method to assess the galileo neo electronic images on (b)(4) 2014.The well images for the unexpected crossmatch compatible wells all appeared as unexpected negative with associated tightly defined red blood cell buttons.Immucor laboratory tested retention products on (b)(4) 2014 and all retention products performed as expected.The customer returned blood samples to immucor for testing, and on (b)(4) 2014 the immucor laboratory duplicated the unexpected compatible crossmatch results.The unexpected crossmatch compatible results have a potential for a transfusion error.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPTURE-R SELECT
Type of Device
SOLID PHASE SYSTEM FOR THE IMMOBILIZATION OF HUMAN ERYTHROCYTES
Manufacturer (Section D)
IMMUCOR, INC.
3130 gateway drive
norcross GA 30091 562
Manufacturer (Section G)
IMMUCOR, INC.
3130 gateway drive
norcross GA 30091 562
Manufacturer Contact
howard yorek
3130 gateway drive
norcross, GA 30091-5625
7704412051
MDR Report Key3734631
MDR Text Key20696708
Report Number1034569-2014-00057
Device Sequence Number1
Product Code KSS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
BK040024
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Medical Technologist
Type of Report Initial
Report Date 04/07/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Medical Technologist
Device Expiration Date01/25/2015
Device Catalogue Number0006446
Device Lot NumberSC262
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/14/2014
Initial Date FDA Received04/08/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/25/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age88 YR
Patient Weight66
-
-