(b)(4).Method: the complaint mr290v vented autofeed humidification chamber was returned to fisher & paykel healthcare in (b)(4) and was visually inspected.Results: visual inspection revealed there was a break in the water feedset tube at the point of insertion to the water bag spike.The surface at the break was rough (not smoothly cut).A lot check revealed one other similar complaint of this nature for lot number 130731.Conclusion: the damage appears to be the result of the tube being pulled away from the spike, possibly due to the feedset being caught or under tension.Past investigations have shown that the problem has been caused by the user removing the spike by grasping the tubing instead of the spike.We have conducted extensive testing of the mr290 chamber, with particular emphasis on feedset breaks.Significantly we have not been able to replicate failure of the feedset tube at the chamber dome in any of our testing.The specification for the chamber requires that the feedset tube should have a breaking strain of (b)(4).During production, pull testing of the feedset strength at both spike and dome end is performed (b)(4) on mr290 chambers from each production line.If any faults are detected the whole batch is placed on hold for investigation.Additionally all chambers are pressure tested before they leave the production line and any holes or leaks in the feedset are identified during this process.Chambers that fail any of these tests are discarded.The mr290 was used in conjunction with a fisher & paykel healthcare rt202 adult breathing circuit which user instructions state: "perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient." "set appropriate ventilator alarms." (b)(4).
|