PHYSIO-CONTROL, INC LIFEPAK(R) 20 DEFIBRILLATOR/MONITOR; DC-DEFIBRILLATOR, LOW-ENERGY, (INCLUDING PADDLES) Back to Search Results
Model Number 20
Device Problem Aborted Charge (2288)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/12/2014
Event Type  malfunction  
Event Description
The customer reported that during a patient event, while charging defibrillation energy for delivery, the device removed the charge.The customer then successfully delivered 200 and 300 joule shocks to the patient successfully.The patient did not suffer any adverse effects as a result of the reported failure.The customer stated that the device removed the charge when the therapy cable was moved around near the therapy connector portion of the cable.
 
Manufacturer Narrative
(b)(4).Physio-control evaluated the device and verified the reported failure.Physio replaced the therapy connector assembly and observed proper device operation through functional and performance testing.The device was then returned to the customer for use.
 
Manufacturer Narrative
Physio-control further evaluated the removed therapy connector assembly.Physio-control observed that the low voltage contact of the therapy connector assembly was loose and physically damaged.The loose contact would cause an open circuit when the therapy cable assembly was manipulated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) 20 DEFIBRILLATOR/MONITOR
Type of Device
DC-DEFIBRILLATOR, LOW-ENERGY, (INCLUDING PADDLES)
Manufacturer (Section D)
PHYSIO-CONTROL, INC
11811 willows road ne
po box 97006
redmond WA 98073 970
Manufacturer (Section G)
PHYSIO-CONTROL, INC
11811 willows road ne
po box 97006
redmond WA 98073 970
Manufacturer Contact
jason march
4258674000
MDR Report Key3736024
MDR Text Key4276801
Report Number3015876-2014-00399
Device Sequence Number1
Product Code LDD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K063119
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 03/13/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number20
Device Catalogue Number3202488
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/14/2014
Is the Reporter a Health Professional? Yes
Device Age10 YR
Event Location Hospital
Initial Date Manufacturer Received 05/27/2014
Initial Date FDA Received04/09/2014
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received06/09/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/27/2004
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-