MAUDE Adverse Event Report: MEDTRONIC MED REL MEDTRONIC PUERTO RICO MAXIMO II CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MED REL MEDTRONIC PUERTO RICO MAXIMO II CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO Back to Search Results
Model Number D284TRK
Device Problems Premature Discharge of Battery (1057); Output above Specifications (1432)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  Injury  
Event Description
It was reported that the device had early battery depletion from high lv (left ventricular) outputs.It was noted that the left ventricular (lv) lead required the high outputs due to patient anatomy.The device was explanted and replaced.The lead remains in use.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
The information submitted reflects all relevant data received.If additional relevant information is received, a supplemental report will be submitted.Concomitant products: 419678, implantable pacing lead, implanted: (b)(6) 2011; a 407645, implantable pacing lea,d implanted: (b)(6) 2011.(b)(4).
 
Manufacturer Narrative
Product event summary: the device was returned and analyzed.Analysis of the device revealed normal battery depletion.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAXIMO II CRT-D
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
Manufacturer (Section D)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
ceiba norte industrial park, r
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC CARDIAC RHYTHM DISEASE MGMT
8200 coral sea st ne
mounds view MN 55112
Manufacturer Contact
nashoane fulwood-kelley
8200 coral sea st ne
mounds view, MN 55112
7635260583
MDR Report Key3745258
MDR Text Key4226234
Report Number3004209178-2014-06830
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative,company representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 02/07/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date03/14/2012
Device Model NumberD284TRK
Device Catalogue NumberD284TRK
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/12/2014
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/10/2014
Initial Date FDA Received04/11/2014
Supplement Dates Manufacturer ReceivedNot provided
04/10/2014
Supplement Dates FDA Received06/10/2014
09/14/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/27/2010
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
694758 IMPLANTABLE TACHY LEAD
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age00072 YR
-
-