SIEMENS HEALTHCARE DIAGNOSTICS INC. ADVIA CENTAUR XP; CLINICAL CHEMISTRY ANALYZER Back to Search Results
Model Number ADVIA CENTAURXP
Device Problem False Positive Result (1227)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/20/2014
Event Type  malfunction  
Event Description
Discordant, (b)(6) results above the cutoff for mandatory confirmation testing were obtained on two patient samples on an advia centaur xp instrument.The discordant results were not reported to the physician(s).The samples were repeated on the same instrument and resulted (b)(6).It is unknown if the corrected results were reported to the physician(s).There are no reports of patient intervention or adverse health consequences due to the discordant, (b)(6) results.
 
Manufacturer Narrative
A siemens customer service engineer (cse) was dispatched to the customer site.After evaluation of the instrument and instrument data, the cse discovered that the acid pump was damaged and replaced it.The cse also replaced the probe wash aspirate and base pump.The cause of the discordant, (b)(6) results is unknown.The instrument is performing according to specifications.No further evaluation of the device is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVIA CENTAUR XP
Type of Device
CLINICAL CHEMISTRY ANALYZER
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS INC.
511 benedict avenue
tarrytown NY 10591
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS
manufacturing limited
registration number: 8020888
chapel lane, swords, co., dublin
EI  
Manufacturer Contact
cassandra kocsis
511 benedict avenue
tarrytown 10591
9145242687
MDR Report Key3749342
MDR Text Key4547174
Report Number2432235-2014-00285
Device Sequence Number1
Product Code MOI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K971418
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Health Professional
Type of Report Initial
Report Date 03/20/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberADVIA CENTAURXP
Device Catalogue Number078-A010-01
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/20/2014
Initial Date FDA Received04/14/2014
Was Device Evaluated by Manufacturer? No
Date Device Manufactured04/23/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-