Model Number MODEL 250
Device Problems Use of Incorrect Control/Treatment Settings (1126); Device Inoperable (1663); Improper or Incorrect Procedure or Method (2017)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/04/2014
Event Type  malfunction  
Event Description
During review of the in-house programming/diagnostic history database, it was observed that on office visit on (b)(6) 2014 the patient's settings were different than what were programmed at the previous office visit on (b)(6) 2014.The settings found were indicative of a faulted diagnostic test which occurred on (b)(6) 2014.The device was not interrogated prior to the patient leaving the office on (b)(6) 2014 as recommended by device manufacturer to ensure the device is at the correct settings; therefore, the settings were not corrected prior to the patient leaving the office.The settings were corrected on (b)(6) 2014.No patient adverse events were reported.
 
Manufacturer Narrative
Analysis of programming history.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING SOFTWARE
Manufacturer (Section D)
CYBERONICS INC
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS INC
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key3752090
MDR Text Key19197798
Report Number1644487-2014-01015
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial
Report Date 03/17/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 250
Device Lot Number1063524
Other Device ID NumberVERSION 8.1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Event Location Other
Initial Date Manufacturer Received 03/17/2014
Initial Date FDA Received04/15/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age29 YR
-
-