It was reported that during a da vinci surgical procedure, the pk dissecting forceps instrument would not coagulate.Nothing reportedly fell into the patient.The planned surgical procedure was completed and no patient harm, adverse outcome or injury was reported.The customer was not sure of the date of the surgical procedure, but the instrument was given to him on (b)(4) 2014.
|
The instrument was returned and evaluated.Failure analysis was unable to confirm the reported complaint.The returned instrument was connected to a pk generator and passed a cautery test.The instrument passed electrical continuity testing.Failure analysis investigation found various scratch marks at the distal end of the main tube exhibiting light material removal.The scratches were approximately.031 -.163 in length and not axially aligned with the tube.No other damage was found.The customer reported complaint does not itself constitute a mdr reportable event; however, the tube abrasions with material removed found during failure analysis investigation, could cause or contribute to an adverse event, if to recur.
|