MAUDE Adverse Event Report: PHYSIO-CONTROL, INC LIFEPAK(R) 20 DEFIBRILLATOR/MONITOR; DC-DEFIBRILLATOR, LOW-ENERGY, (INCLUDING PADDLES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHYSIO-CONTROL, INC LIFEPAK(R) 20 DEFIBRILLATOR/MONITOR; DC-DEFIBRILLATOR, LOW-ENERGY, (INCLUDING PADDLES) Back to Search Results
Model Number 20
Device Problem Inappropriate/Inadequate Shock/Stimulation (1574)
Patient Problem No Patient Involvement (2645)
Event Date 03/03/2014
Event Type  malfunction  
Event Description
The customer reported that the device was displaying the service indicator.There was no report of patient use associated with the reported issue.Upon evaluation of the device, physio observed a failure that resulted in a partial loss of defibrillator output energy due to a loss of the negative portion of the biphasic output waveform.The defibrillator output energy could be reduced by up to approximately 20% from the selected energy level.
 
Manufacturer Narrative
(b)(4): physio-control examined the customer's device and verified the reported failure.Physio then replaced the therapy pcb assembly and after observing proper device operation through functional and performance testing the unit was returned to the customer for use.Physio further examined the removed therapy pcb assembly and determined that the cause of the reported failure was that a filter, designator fl29, had broken off of the assembly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) 20 DEFIBRILLATOR/MONITOR
Type of Device
DC-DEFIBRILLATOR, LOW-ENERGY, (INCLUDING PADDLES)
Manufacturer (Section D)
PHYSIO-CONTROL, INC
11811 willows road ne
po box 97006
redmond WA 98073 970
Manufacturer (Section G)
PHYSIO-CONTROL, INC
11811 willows road ne
po box 97006
redmond WA 98073 970
Manufacturer Contact
jason march
4258674000
MDR Report Key3752233
MDR Text Key4466375
Report Number3015876-2014-00424
Device Sequence Number1
Product Code LDD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K063119
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 03/19/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number20
Device Catalogue Number3202487
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/06/2014
Is the Reporter a Health Professional? Yes
Device Age11 YR
Event Location Hospital
Initial Date Manufacturer Received 03/19/2014
Initial Date FDA Received04/15/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/26/2003
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-