Model Number 9800
Device Problems Erratic or Intermittent Display (1182); Device Stops Intermittently (1599)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/30/2014
Event Type  malfunction  
Event Description
The customer reported the fluoroscopic image was intermittently lost from both monitors.This issue will effectively eliminate the ability to view a real time image.No pt death or serious injury was reported related to this event.
 
Manufacturer Narrative
A ge service representative performed an on site investigation.The reported issue could not be duplicated.The boards and connectors were reseated during the service call.The related power supplies were also evaluated and confirmed to be in specification.The system was tested and found to be working as intended and put back into service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
9800
Type of Device
FLUOROSCOPIC X-RAY
Manufacturer (Section D)
GE OEC MEDICAL SYSTEMS (SLC)
384 wright brothers dr.
salt lake city UT 84116
Manufacturer (Section G)
GE OEC MEDICAL SYSTEMS (SLC)
384 wright brothers dr.
salt lake city UT 84116
Manufacturer Contact
384 wright brothers dr.
salt lake city, UT 84116
8015364952
MDR Report Key3755607
MDR Text Key4483859
Report Number1720753-2014-01434
Device Sequence Number1
Product Code JAA
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 02/12/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9800
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/30/2014
Initial Date FDA Received02/12/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/01/2004
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-