Catalog Number 0606600000 |
Device Problem
Disassembly (1168)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 03/27/2014 |
Event Type
malfunction
|
Event Description
|
It was reported that following a procedure, the revolution cement gun was found to have a screw missing on the end.The procedure was completed successfully.No delay, no medical intervention and no adverse consequences were reported with this event.
|
|
Event Description
|
It was reported that following a procedure the revolution cement gun was found to have a screw missing on the end.The procedure was completed successfully.No delay, no medical intervention and no adverse consequences were reported with this event.
|
|
Manufacturer Narrative
|
A follow up report will be filed after the quality investigation has been completed.
|
|
Manufacturer Narrative
|
One of the two screws at the distal end of the cradle was confirmed to be missing on the returned device.The device was repaired and returned to the user facility.
|
|