The patient rec'd a second degree burn mark on the arm, after the tc sensor was removed.The temperature was set to 43 degrees celsius, and the electrode was placed on the same location for 60 minutes.According to the doctor, the patient was going through a septic shock, and this can lead to decreased blood circulation.
|
According to the complaint, the temperature of the sensor was set to 43 degrees celsius and the sensor was relocated every 60 minutes.This is within the operator's manual for patients with normal blood circulation.However, in the operators manual there is a warning about always paying attention to the use of the sensors on patients in shock due to risk of skin damage.Warning - risk of skin damage.Long-term hyperthermia may blister skin.When producing local hyperemia by means of hyperthermia, a certain risk of applying temperatures harmful to the skin is always present, although the risk is limited due to the control system of the instrument.Always pay attention to the use of hyperthermia for special patients - e.G.Patients in shock, patients with low blood pressure, and patients with vascular constrictions.The placement of the sensor for 60 minutes could explain the burn mark on this patient.To analyze the case further the sensor will be investigated to see if a product malfunction also contributed to this case.
|