• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMERICAN MEDICAL SYSTEMS (MN) AMS ACTICON NEOSPHINCTER; IMPLANTED FECAL INCONTINENCE DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMERICAN MEDICAL SYSTEMS (MN) AMS ACTICON NEOSPHINCTER; IMPLANTED FECAL INCONTINENCE DEVICE Back to Search Results
Catalog Number SEE H10
Device Problem Leak/Splash (1354)
Patient Problem Failure of Implant (1924)
Event Date 03/14/2014
Event Type  Injury  
Event Description
It was reported that the patient had their action bowel sphincter removed due to malfunction, hole in system, and that the pump stopped working.No patient complications were reported.
 
Manufacturer Narrative
Cuff catalog # 72401984, serial # (b)(4), expiration date 03/31/2003; pump catalog # 72402287, serial # (b)(4), expiration date 03/01/2006; balloon catalog # 72402105, serial # (b)(4), expiration date 03/05/2006.The device was returned and was visually inspected and functionally tested.The pump and cuff both performed within specification.The balloon had a leak identified that appears to be the result of a sharp instrument which could have occurred during device removal.Cuff manufacture date - 3/1998, pump & balloon manufacture date 3/2001.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS ACTICON NEOSPHINCTER
Type of Device
IMPLANTED FECAL INCONTINENCE DEVICE
Manufacturer (Section D)
AMERICAN MEDICAL SYSTEMS (MN)
10700 bren road w
minnetonka MN 55343
Manufacturer (Section G)
AMERICAN MEDICAL SYSTEMS (MN)
10700 bren road w
minnetonka MN 55343
Manufacturer Contact
sharon zurn
10700 bren road w
minnetonka, MN 55343
9529306000
MDR Report Key3766484
MDR Text Key20786885
Report Number2183959-2014-00129
Device Sequence Number1
Product Code MIP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010020
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 04/08/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue NumberSEE H10
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/08/2014
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/08/2014
Initial Date FDA Received04/23/2014
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age92 YR
-
-