The customer reported pressure/vacuum background check, hemoglobin (hgb) high, and high vacuum low system errors involving the coulter® hmx hematology analyzer with autoloader.The customer performed system diagnostics with beckman coulter customer technical support (cts) via the telephone but was unsuccessful.There were no erroneous patient results generated at the time of the event.Patient care was not impacted.A beckman coulter field service engineer (fse) was dispatched to assess the instrument.
|
Based on root cause, the suspect product was identified to be the reagent(s) used with the instrument.Subsequent products are provided here; it is unknown which lots were in use at the time of the event.Therefore, expiration date and device manufacturing date are also unknown for this event.Brand name: coulter isoton iii diluent; catalog #: 8546733; lot #: 50793f through 50818f.Brand name: coulter isoton 4 diluent; catalog #: 8547148; lot #: 18206f.Brand name: coulter lh series diluent; catalog #: 8547194; lot #: 510409f through 510657f; m405237 through m503553.The manufacturing site address has been updated to correlate to the product.New information about the root cause was completed on 8/12/2015.Root cause changed the 510k to exempt to correlate to the product.(b)(4).A recall was completed and an important product notice letter was sent to customers on 09/09/2015.The root cause has been identified as lot to lot variation in the sodium sulfate used in the reagent, which resulted in a compromised white blood cell (wbc) bath in this event.In addition, the investigation found that the root cause identified would not cause or contribute.
|