• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: KERR CORPORATION SONICFILL COMPOSITE; TOOTH SHADE RESIN MATERIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KERR CORPORATION SONICFILL COMPOSITE; TOOTH SHADE RESIN MATERIAL Back to Search Results
Device Problems Device Operates Differently Than Expected (2913); Material Integrity Problem (2978)
Patient Problem No Code Available (3191)
Event Type  Injury  
Event Description
A doctor's office alleged that two (2) patients had experienced voids within their restorations after placement with sonicfill composite.This is the second of two (2) reports.
 
Manufacturer Narrative
Patient specific with regard to gender, age, and weight was not provided.Although the doctor provided three (3) different shades associated with the voids, he could not verify which lot was used on each patient; therefore, no lot numbers were identified.The lots involved in the alleged incidents include lot numbers 4931574, 4931579, and 4931576.The doctor could not recall patient or incident detail.To date, the patient is doing fine.The doctor drilled out the restoration and redid the procedure using the same product, without further incident.An 'extrusion' and 'void concentration' tests of the returned product was performed, yielding results within specifications.A dhr review revealed that there were no deviations from the manufacturing process.In addition, no similar complaints were received with regard to these lots.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SONICFILL COMPOSITE
Type of Device
TOOTH SHADE RESIN MATERIAL
Manufacturer (Section D)
KERR CORPORATION
1717 west collins avenue
orange CA 92867
Manufacturer (Section G)
KERR CORPORATION
1717 west collins avenue
orange CA 92867
Manufacturer Contact
kerri casino
1717 west collins avenue
orange, CA 92867
7145167634
MDR Report Key3786775
MDR Text Key4443626
Report Number2024312-2014-00276
Device Sequence Number1
Product Code EBF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K091023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Dentist
Type of Report Initial
Report Date 04/03/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Dentist
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/03/2014
Initial Date FDA Received05/02/2014
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other; Required Intervention;
-
-