Our examination revealed that one of the terminals had been broken near a thermostat, which had compromised our double-insulated design.We also found several other internal components that were damaged, indicating that the customer misuse the pad by applying an excessive, localized force on or near the thermostat terminal.We concluded that this report was attributable to misuse by the user, and failure ot follow instructions as to the care and handling of the unit.After investigation qx noted the following: pad appears to have been used in a roller bed.Outline of hw and lead on cover, front and back.Pad bunches, bent/broken lead, burn - harness.Although warning labels and instruction booklets warn against sitting or laying on the product, we have a corrective action project (b)(4) underway to make the design more robust.
|