MAUDE Adverse Event Report: BATTLE CREEK EQUIPMENT THERMOPHORE AUTOMATIC HEAT PACK; ELECTRIC MOIST HEAT PACK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BATTLE CREEK EQUIPMENT THERMOPHORE AUTOMATIC HEAT PACK; ELECTRIC MOIST HEAT PACK Back to Search Results
Model Number 055
Device Problem Thermal Decomposition of Device (1071)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/17/2014
Event Type  malfunction  
Event Description
Customer recently noticed a burn mark on the outer protective sheet they put over the pad.Took off all of the coverings and the actual pad had burned through everything and even onto their therapy bed.
 
Manufacturer Narrative
Our examination revealed that one of the terminals had been broken near a thermostat, which had compromised our double-insulated design.We also found several other internal components that were damaged, indicating that the customer misuse the pad by applying an excessive, localized force on or near the thermostat terminal.We concluded that this report was attributable to misuse by the user, and failure ot follow instructions as to the care and handling of the unit.After investigation qx noted the following: pad appears to have been used in a roller bed.Outline of hw and lead on cover, front and back.Pad bunches, bent/broken lead, burn - harness.Although warning labels and instruction booklets warn against sitting or laying on the product, we have a corrective action project (b)(4) underway to make the design more robust.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THERMOPHORE AUTOMATIC HEAT PACK
Type of Device
ELECTRIC MOIST HEAT PACK
Manufacturer (Section D)
BATTLE CREEK EQUIPMENT
freemont IN
Manufacturer Contact
randy newsome
702 south reed street
freemont, IN 46737
2699626181
MDR Report Key3788434
MDR Text Key20265779
Report Number1811605-2014-00084
Device Sequence Number1
Product Code IRT
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,User Facility
Reporter Occupation Not Applicable
Type of Report Initial
Report Date 04/01/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number055
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/24/2014
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/24/2014
Initial Date FDA Received04/01/2014
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-