Model Number INSTATRAK 3500
Device Problem Device Inoperable (1663)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/27/2013
Event Type  malfunction  
Event Description
The customer reported that the system locked up.This resulted in a total loss of navigation functionality.There is no report of injury or death associated with this event.
 
Manufacturer Narrative
A ge representative contacted the customer by phone and directed them in repairing the system.The customer performed a database recovery, and removed patient data from previous years from the system.It was recommended that the customer keep data on the system for six months.The system operates as intended and was returned to service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSTATRAK 3500
Type of Device
RADIOLOGICAL IMAGE PROCESSING
Manufacturer (Section D)
GE OEC MEDICAL SYSTEMS (SLC)
384 wright brothers dr.
salt lake city UT 84116
Manufacturer (Section G)
GE OEC MEDICAL SYSTEMS (SLC)
384 wright brothers dr.
salt lake city UT 84116
Manufacturer Contact
384 wright brothers dr.
salt lake city, UT 84116
8015364952
MDR Report Key3792309
MDR Text Key4529123
Report Number1720753-2014-00386
Device Sequence Number1
Product Code LLZ
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 01/10/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberINSTATRAK 3500
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/27/2013
Initial Date FDA Received01/10/2014
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-