Catalog Number 20709
Device Problems Occlusion Within Device (1423); Failure to Sense (1559)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/01/2011
Event Type  malfunction  
Event Description
The customer contact reported that during preventive maintenance testing at the user facility, the device did not alarm when a distal occlusion was present.There were no reported adverse patient events or delays in critical therapy while the device was in clinical use.Though requested, no additional information was provided.
 
Manufacturer Narrative
At this time the customer will not be returning the device for evaluation.The device was not returned to hospira for testing and investigation; therefore, attribution of the issue to the device could not be determined.This report is being made as part of a retrospective review.No further testing or investigation will be performed.This report represents all the information known by the reporter upon query by hospira personnel.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LC PCA NEW 802.11 AB
Type of Device
80MEA
Manufacturer (Section D)
HOSPIRA COSTA RICA LTD.
zona franca global
la aurora heredia
CS 
Manufacturer (Section G)
HOSPIRA COSTA RICA LTD.
zona franca global
la aurora heredia
CS  
Manufacturer Contact
juergen schmider, md, vp
275 n. field dr.
bldg. no. h2-1east, dept no. 097u
lake forest, IL 60045
2242125740
MDR Report Key3806752
MDR Text Key4477969
Report Number9615050-2014-00482
Device Sequence Number1
Product Code MEA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K042800
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Type of Report Initial
Report Date 12/13/2011
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number20709
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/13/2011
Initial Date FDA Received01/22/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/01/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-