• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ASCENSION ORTHOPEDICS MGT MOVEMENT PROX. PHAL. TOTAL SZ 20; MOVEMENT GREAT TOE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ASCENSION ORTHOPEDICS MGT MOVEMENT PROX. PHAL. TOTAL SZ 20; MOVEMENT GREAT TOE Back to Search Results
Lot Number 10-1204
Device Problem Peeled/Delaminated (1454)
Patient Problems Pain (1994); Discomfort (2330)
Event Type  Injury  
Event Description
This is the second report of two reports concerning the same patient.(see mfg report number 1651501-2014-00021).This report concerns the movement great toes system (mgt) movement proximal phalanx.Total size 20 (product id mgt-890-20ppt).It was reported: 'surgery was an mgt total toe.Surgery went fine but the patient soon began to have pain.The surgeon went back in to replace the implant and found that the cobalt chrome had peeled off the head of the implant.' additional information was received from the distributor on (b)(6)-2014.The date of the initial implant surgery was (b)(6)-2013.The implant was implanted in the left foot of the patient.The patient had discomfort soon after the surgery which progressed until the implant was removed (and toe was later fused) on (b)(6)-2014.The patient condition to date was reported as 'improved'.The implant was reported given to the patient and is not available for evaluation.Additional information was requested by integra.
 
Manufacturer Narrative
The device involved in the reported incident is not available for evaluation.An investigation has been initiated based on the reported information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MGT MOVEMENT PROX. PHAL. TOTAL SZ 20
Type of Device
MOVEMENT GREAT TOE
Manufacturer (Section D)
ASCENSION ORTHOPEDICS
austin TX
Manufacturer Contact
caren finkelstein
315 enterprise drive
plainsboro, NJ 08536-3832
6099362341
MDR Report Key3810009
MDR Text Key4523722
Report Number1651501-2014-00022
Device Sequence Number1
Product Code LZJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102549
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Distributor
Reporter Occupation Other
Type of Report Initial
Report Date 04/07/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Lot Number10-1204
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/07/2014
Initial Date FDA Received05/01/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age68 YR
-
-