• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS INC STREAMLAB ANALYTICAL WORKCELL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS INC STREAMLAB ANALYTICAL WORKCELL Back to Search Results
Model Number STREAMLAB ANALYTICAL WORKCELL
Device Problem Device Emits Odor (1425)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/19/2014
Event Type  malfunction  
Event Description
The customer reported seeing smoke coming from the streamlab centrifuge module.The customer powered off the device.No one was injured or treated for smoke inhalation.There were no reports of adverse health consequences due to the smoke emitted from the device.
 
Manufacturer Narrative
A siemens customer service engineer (cse) was dispatched to the customer site.After evaluation of the device, the cse determined that the cause of the smoke emitted was a compressor burnt out due to a leak in the system.The cse replaced and recharged the compressor.The instrument is performing within specifications.No further evaluation of the device is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STREAMLAB ANALYTICAL WORKCELL
Type of Device
STREAMLAB ANALYTICAL WORKCELL
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS INC
101 silvermine road
brookfield CT 06804
Manufacturer (Section G)
INPECO S.P. AVIA G. DI VITTORIO
registration #: 3005509212
via g. di vittorio
segrate, 11 I -200
IT   11 I-2009
Manufacturer Contact
cassandra kocsis
511 benedict ave
tarrytown, NY 10591
9145242687
MDR Report Key3811798
MDR Text Key4474467
Report Number1226181-2014-00268
Device Sequence Number1
Product Code LXG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Health Professional
Type of Report Initial
Report Date 04/19/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSTREAMLAB ANALYTICAL WORKCELL
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/19/2014
Initial Date FDA Received05/15/2014
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-