The pt's weight is unk.The hospital was not able to provide the information.As a conservative measure, an mdr is being submitted because the transition tubing separated from the proximal bond but remained intact to the device.The angiosculpt device was received for evaluation.Visual examination confirmed a bent and damaged scoring element.The distal bond is damaged and the intermediate bond is over the proximal marker band.The rx port is lacerated and the shaft is damaged and twisted.The transition tubing is twisted and separated at the proximal bond but remained intact to the intermediate bond.Based on the lab analysis, it leads to suggest that the device experienced excessive exertion of force applied by the user.According to the instructions for use (ifu) for the angiosculpt catheter, retained device components is listed as a possible adverse effect of the procedure.
|