• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS, INC. 1818910 PINN CAN BONE SCREW 6.5MMX25MM; HIP OTHER IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS, INC. 1818910 PINN CAN BONE SCREW 6.5MMX25MM; HIP OTHER IMPLANT Back to Search Results
Catalog Number 121725500
Device Problems Loose or Intermittent Connection (1371); Loss of Osseointegration (2408); Osseointegration Problem (3003)
Patient Problems Host-Tissue Reaction (1297); Pain (1994); Tissue Damage (2104); Discomfort (2330); Osteolysis (2377); Ambulation Difficulties (2544); Inadequate Osseointegration (2646); Limited Mobility Of The Implanted Joint (2671); Not Applicable (3189)
Event Date 04/25/2014
Event Type  Injury  
Event Description
The patient was revised because of a loose cup with some osteolysis.
 
Manufacturer Narrative
Additional narrative: this complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
The devices associated with this report were not returned.A review of the device history records for the c68ek4000 lot code did not reveal any related manufactuirng deviations or anomalies.A complaint database search finds no other reported incidents against the remaining product and lot combinations.Requests for additional investigational inputs were made in accordance with wi-7915 appendix a.No additional information was obtained.Medical records were obtained and reviewed by a medical professional.With the limited amount of information provided, it cannot be determined that the complaint is product related.The investigation could not verify or identify any product contribution to the reported event with the information provided.Based on the inability to identify root cause, the need for corrective action was not indicated.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
 
Manufacturer Narrative
The patient was revised because of a loose cup with some osteolysis.(b)(4).The devices associated with this report were not returned.A review of the device history records for the c68ek4000 lot code did not reveal any related manufacturing deviations or anomalies.A complaint database search finds no other reported incidents against the remaining product and lot combinations.Requests for additional investigational inputs were made in accordance with wi-7915 appendix a.No additional information was obtained.The investigation could not verify or identify any product contribution to the reported event with the information provided.Based on the inability to identify root cause, the need for corrective action was not indicated.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
 
Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.
 
Event Description
Update rec'd 7/25/2014- pfs and medical records received.After review of the medical records there is no new additional information that would affect the existing mdr decision.
 
Manufacturer Narrative
Additional info: update rec'd 7/23/2014 - litigation papers received.In addition to what was previously reported, litigation alleges the patient suffers from pain, discomfort, and decreased mobility.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Update ad (b)(6) 2018 ppf, pfs and medical records received.In addition to what were previously alleged, pfs also alleges walking difficulty and difficulty carrying out normal daily activities.Ppf has no new allegation.After review of medical records for mdr reportability, it was stated that the patient was revised to address loosening of the cup and elevated metal ion levels.Revision notes reported a small cyst, 15 cc of brown cloudy fluid consistent with metallosis, osteolytic lesion, two screws were removed easily, and metal debris throughout the adjacent soft tissue.Added stem due to the reported elevated metal ion levels but was not revised.There were no reported lab results.Doi: (b)(6) 2009; dor: (b)(6) 2014; left hip.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Corrected:h6(device codes) product complaint # : (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : null, device history batch : null, device history review : null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINN CAN BONE SCREW 6.5MMX25MM
Type of Device
HIP OTHER IMPLANT
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic drive
warsaw IN 46582 0988
MDR Report Key3816622
MDR Text Key4395452
Report Number1818910-2014-18974
Device Sequence Number1
Product Code NDJ
Combination Product (y/n)N
PMA/PMN Number
PK983014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Distributor,consumer,distributor,he
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 07/25/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date06/06/2019
Device Catalogue Number121725500
Device Lot NumberDW7BL4000
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 02/22/2015
Initial Date FDA Received05/19/2014
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Not provided
Not provided
04/13/2018
09/03/2019
Supplement Dates FDA Received08/04/2014
08/06/2014
08/08/2014
03/04/2015
05/10/2018
09/09/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age48 YR
Patient Weight147
-
-