Device 1 of 2.Reference mfr report # 1627487-2014-12348.It was reported the pt post stimulation.An sjm rep interrogated the pt's system one week post implant procedure.All of the contacts had invalid impedance on one of the leads.An x-ray revealed the lead pulled out of the ipg header.The pt's lead revision procedure was (b)(6) 2014.The physician reattached to leads to the ipg but there were impedance issues.The physician noticed fluid in the ipg header.The ipg was explanted and replaced.Extensions were attached to the existing leads then connected to the new ipg.The pt's stimulation was restored.Note the pt received two leads, occipital placement (off label use), from the same lot number.
|