Catalog Number 6500000000
Device Problem Component Missing (2306)
Patient Problem No Patient Involvement (2645)
Event Date 04/14/2014
Event Type  Injury  
Event Description
It was reported by the customer that they allegedly dropped a patient by missing a safety hook.The customer reported that the ambulance was parked on a grade at the hospital with a patient on the cot and as they came out of the back of the truck, the wheels swiveled and they missed the hook.There was patient involvement; however, no adverse consequence or clinically relevant delay in treatment was reported.
 
Event Description
It was reported by the customer that they allegedly dropped a patient by missing a safety hook.The customer reported that the ambulance was parked on a grade at the hospital with a patient on the cot and as they came out of the back of the truck, the wheels swiveled and they missed the hook.The patient received scrapes that required medical intervention from the event.
 
Manufacturer Narrative
A visual and functional inspection was performed by a stryker field service representative.It was alleged that the cot missed the safety bar and dropped out of the ambulance with a patient.It was further reported that the ambulance was parked on an incline (left to right) which caused the cot and the patient to tip onto one load wheel high enough for the safety bar to miss the hook.The patient received scrapes that required medical intervention from the event.In addition, it was reported that there was only one user at the foot end during the unloading procedure.Upon evaluation, the field technician was unable to duplicate the alleged event and found no defects with the device.The issue was resolved for the customer by instructing the customer on how to properly unload the cot.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWER PRO AMBULANCE COT - OBS
Type of Device
STRETCHER, WHEELED
Manufacturer (Section D)
STRYKER MEDICAL-KALAMAZOO
3800 east centre avenue
portage MI 49002
Manufacturer (Section G)
STRYKER MEDICAL-KALAMAZOO
3800 east centre avenue
portage MI 49002
Manufacturer Contact
anthony noora
3800 east centre avenue
portage, MI 49002
2693292100
MDR Report Key3821328
MDR Text Key4398660
Report Number0001831750-2014-02992
Device Sequence Number1
Product Code FPO
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/23/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number6500000000
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/23/2014
Initial Date FDA Received05/20/2014
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received06/04/2014
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-