• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BATTLE CREEK EQUIPMENT THERMOPHORE AUTOMATIC HEAT PACK; ELECTRIC MOIST HEAT PACK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BATTLE CREEK EQUIPMENT THERMOPHORE AUTOMATIC HEAT PACK; ELECTRIC MOIST HEAT PACK Back to Search Results
Model Number 055
Device Problems Hole In Material (1293); Sparking (2595)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/07/2014
Event Type  malfunction  
Event Description
Customer called and stated the pad had sparked by a hole in it.
 
Manufacturer Narrative
Our examination revealed that one of the terminals had been broken near a thermostat, which had compromised our double-insulated design.We also found several other internal components that were damaged, indicating that the customer misused the pad by applying an excessive, localized force on or near the thermostat terminal.We concluded that this report was attributable to misuse by the user, and failure to following instructions as to the care and handling of the unit.Although warning labels and instruction booklets warn against sitting or laying on the product, we have a corrective action project (b)(4) underway to make the design more robust.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THERMOPHORE AUTOMATIC HEAT PACK
Type of Device
ELECTRIC MOIST HEAT PACK
Manufacturer (Section D)
BATTLE CREEK EQUIPMENT
fremont IN
Manufacturer Contact
randy newsome
702 south reed st
fremont, IN 46737
2699626181
MDR Report Key3823711
MDR Text Key4390568
Report Number1811605-2014-00105
Device Sequence Number1
Product Code IRT
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,User Facility
Reporter Occupation Not Applicable
Type of Report Initial
Report Date 05/01/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number055
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/17/2014
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/17/2014
Initial Date FDA Received05/02/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/01/2002
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-