Product analysis was completed on the lead and the generator.In the pa lab, the device output signal was monitored for more than 24-hrs, while the generator was placed in a simulated body temperature environment.Results showed no signs of variation in the pulse generator¿s output signal and demonstrated that the device provided the expected level of output current for the entire monitoring period.The pulse generator diagnostics were as expected for the programmed parameters.The generator performed according to functional specifications.During the product analysis there were no anomalies found with the pulse generator.Potential contributing factors to the infection have been considered/evaluated for the generator and none were found to exist in this situation.The dhr shows that the pulse generator passed all specifications before it was released.Note that a large portion of the lead assembly (body) including the electrode section was not returned for analysis and therefore a complete evaluation could not be performed on the entire lead product.The condition of the returned lead portion is consistent with conditions that typically exist following an explant procedure.No obvious anomalies were noted.The setscrew marks found on the connector pin provide evidence that, at one point in time, a good mechanical and electrical connection was present.Continuity checks of the returned lead portion were performed, during the visual analysis, with no discontinuities identified.Based on the findings in the product analysis lab, there is no evidence to suggest an anomaly with the returned portion of the device.Note that since a large portion of the lead assembly (body) including the electrode array section was not returned for analysis, an evaluation and resulting commentary cannot be made on that portion of the lead.
|